Key result
A biomimetic phantom composed of hollow microfibers maintained stable diffusion properties over 4 months (mean ADC 7.53 ± 0.16 × 10(-4) mm2/s) and generated analogous fiber tracking to rat heart.
A newly developed biomimetic phantom successfully simulates diffusion in the heart and remains stable over 4 months, providing a tool for developing and validating cardiac diffusion MRI methods.
Provides stable cardiac DTI phantom model; hypothesis-generating for human translation.
PURPOSE: Diffusion magnetic resonance imaging (MRI) is increasingly used to characterize cardiac tissue microstructure, necessitating the use of physiologically relevant phantoms for methods development. Existing phantoms are generally simplistic and mostly simulate diffusion in the brain. Thus, there is a need for phantoms mimicking diffusion in cardiac tissue. MATERIALS AND METHODS: A biomimetic phantom composed of hollow microfibers generated using co-electrospinning was developed to mimic myocardial diffusion properties and fiber and sheet orientations. Diffusion tensor imaging was carried out at monthly intervals over 4 months at 9.4T. 3D fiber tracking was performed using the phantom and compared with fiber tracking in an ex vivo rat heart. RESULTS: The mean apparent diffusion coefficient and fractional anisotropy of the phantom remained stable over the 4-month period, with mean values of 7.53 ± 0.16 × 10(-4) mm(2) /s and 0.388 ± 0.007, respectively. Fiber tracking of the 1st and 3rd eigenvectors generated analogous results to the fiber and sheet-normal direction respectively, found in the left ventricular myocardium. CONCLUSION: A biomimetic phantom simulating diffusion in the heart was designed and built. This could aid development and validation of novel diffusion MRI methods for investigating cardiac microstructure, decrease the number of animals and patients needed for methods development, and improve quality control in longitudinal and multicenter cardiac diffusion MRI studies.
No takes yet. Share an insight, caveat, or question.
Teh et al. (2015) studied Cardiac tissue microstructure simulation. Biomimetic phantom composed of hollow microfibers vs. Ex vivo rat heart was evaluated on Mean apparent diffusion coefficient and fractional anisotropy. A biomimetic phantom composed of hollow microfibers maintained stable diffusion properties over 4 months (mean ADC 7.53 ± 0.16 × 10(-4) mm2/s) and generated analogous fiber tracking to rat heart.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: