SENSE can be used in steady-state gradient-echo imaging to increase SNR efficiency by balancing reduced phase encoding with increased repetition time.
Answers are grounded in this paper and cite their sources.
May enhance SNR efficiency in contrast-enhanced MR angiography; leaves open prospective clinical validation before practice change.
Sensitivity encoding (SENSE) with a receiver coil array is typically used as a means of reducing the scan time in MRI. The speed benefit usually comes at some expense in terms of the signal-to-noise ratio (SNR) efficiency, which has been notorious as the main downside of SENSE and parallel MRI in general. In this work it is shown that in steady-state gradient-echo imaging the parallel approach may as well be used to increase the SNR efficiency. The basic idea is to balance reduced phase encoding by increasing the repetition time. In this fashion both the acquisition duty cycle and the steady-state magnetization can be enhanced, resulting in considerable net gains in SNR yield. It is argued that the reduction factor in parallel imaging is essentially an additional degree of freedom in optimizing the SNR. The optimal SENSE factor depends on scan, tissue, and hardware parameters, assuming values up to 3.0 and higher. The achievable SNR benefit also depends on the spoiling regime and is most pronounced for RF-spoiled techniques. The proposed mechanism is demonstrated by simulations and phantom experiments, as well as by contrast-enhanced angiography in vivo, achieving an approximate doubling of the SNR efficiency.
No takes yet. Share an insight, caveat, or question.
Weiger et al. (2004) studied this question.