Observational (n=17)
Answers are grounded in this paper and cite their sources.
PURPOSE: To develop a high isotropic-resolution sequence to evaluate intracranial vessels at 3.0 Tesla (T). MATERIALS AND METHODS: Thirteen healthy volunteers and 4 patients with intracranial stenosis were imaged at 3.0T using 0.5-mm isotropic-resolution three-dimensional (3D) Volumetric ISotropic TSE Acquisition (VISTA; TSE, turbo spin echo), with conventional 2D-TSE for comparison. VISTA was repeated for 6 volunteers and 4 patients at 0.4-mm isotropic-resolution to explore the trade-off between SNR and voxel volume. Wall signal-to-noise-ratio (SNR(wall) ), wall-lumen contrast-to-noise-ratio (CNR(wall-lumen) ), lumen area (LA), wall area (WA), mean wall thickness (MWT), and maximum wall thickness (maxWT) were compared between 3D-VISTA and 2D-TSE sequences, as well as 3D images acquired at both resolutions. Reliability was assessed by intraclass correlations (ICC). RESULTS: Compared with 2D-TSE measurements, 3D-VISTA provided 58% and 74% improvement in SNR(wall) and CNR(wall-lumen) , respectively. LA, WA, MWT and maxWT from 3D and 2D techniques highly correlated (ICCs of 0.96, 0.95, 0.96, and 0.91, respectively). CNR(wall-lumen) using 0.4-mm resolution VISTA decreased by 27%, compared with 0.5-mm VISTA but with reduced partial-volume-based overestimation of wall thickness. Reliability for 3D measurements was good to excellent. CONCLUSION: The 3D-VISTA provides SNR-efficient, highly reliable measurements of intracranial vessels at high isotropic-resolution, enabling broad coverage in a clinically acceptable time.
No takes yet. Share an insight, caveat, or question.
Qiao et al. (2011) conducted an observational in Intracranial stenosis and healthy controls (n=17). 3D Volumetric ISotropic TSE Acquisition (3D-VISTA) MRI at 0.5-mm and 0.4-mm isotropic resolution vs. Conventional 2D-TSE was evaluated on Wall signal-to-noise-ratio (SNRwall), wall-lumen contrast-to-noise-ratio (CNRwall-lumen), lumen area, wall area, mean wall thickness, and maximum wall thickness (ICC 0.91-0.96; 58% SNRwall and 74% CNRwall-lumen improvement).
Synapse has enriched 10 related papers. Consider them for comparative context: