Answers are grounded in this paper and cite their sources.
A new technique has been developed using a tiny thermistor 0.3–0.5 mm in diameter to determine thermal conductivity of biomaterials over a wide temperature range. Based on steady-state spherical heat transfer in an infinite homogeneous medium, thermal conductivity of the measured medium can be determined from the power applied and the temperature rise of the self-heated thermistor. Compared with accepted literature, relative measurement errors of thermal conductivity of standard samples (water and ice, as well as aqueous glycol and CaCl2 solutions) are less than 5% in the temperature range 233–313K. Using the developed technique, thermal conductivities of rabbit liver, kidney, heart, and carotid artery were determined in the temperature range 233–293K.
No takes yet. Share an insight, caveat, or question.
Zhang et al. (2002) studied this question.