Real-time processing of MRIs is reported as a method of 3-D tracking of mechanical devices within the field of view using passive microcoil fiducials. The specific implementation described makes use of two scan planes for full 3-D tracking of a 5-DOF manipulator arm used for prostate biopsy under image guidance. Real-time tracking was observed with a maximum update rate of 0.42 frames per second for a maximum probe velocity of 10 mm/s. The localization of fiducials had a mean error of 0.36 (±0.17) mm ($p ≪ 0.02$), leading to a mean error in the needle tip position of 2.6 (±0.3) mm ($p ≪ 0.05$).