The fabrication and characterization of forward-looking active catheter actuated by titanium-nickel (Ti-Ni) shape memory alloy (SMA) springs are described. The catheter has been designed for wide-range observation of an affected area inside a blood vessel when the blood vessel is occluded. The developed active catheter consists of eight Ti-Ni SMA spring actuators for actuation of catheter tip, an ultrasonic transducer for forward-looking, a guide wire, a polyurethane tube for coating, and spiral wirings for realization of various flexure motions of catheter tip using Ti-Ni SMA actuators. The size of the catheter is 3.5 mm in diameter and 60 mm in length of the sum of transducer and actuator sections. Ti-Ni SMA springs were fabricated from a Ti-50.9at.%Ni sheet by electrochemical etching with a mixed solution of ethanol and lithium chloride. The catheter was assembled by hand under a stereomicroscope. The tip of the produced catheter was able to move in parallel toward at least eight directions by controlling an applied current to Ti-Ni SMA springs. We have confirmed that the active catheter was able to observe an object settled in the front.