Answers are grounded in this paper and cite their sources.
A resonant waveguide grating structure based on periodically perforated metal slits (PPMSs) was specified in experiments for metal-slit cavity resonance in the frequency range of 0.1–1 terahertz (THz), including the structural symmetry, thickness and distributed length of a metal-slit array. In dielectric-medium perturbation experiments, 2D spatial confinement of a metal-slit cavity resonance wave in a large area was characterized with distinct transmittance dip at 0.5 THz and relevant redshift effect of glucose dispersion responsivity. This metal-slit-cavity resonance wave was thus used as an optical probe of one PPMS-based glucose refractometer for recognizing molecular density range of 0–24 µg/mm2, a resolution density of 1.6 µg/mm2, and a trace density of 2.61 µg/mm2. For a megahertz-frequency resolution of THz optomechatronic system, PPMS-based glucose refractometer potentially could realize a detection of limit lower than 2.2 ng/mm2, equaling 1.8 mg/dL, and would also be comparable to other sensing schemes of glucose through optical probes of infrared ray–visible light waves.
No takes yet. Share an insight, caveat, or question.
You et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: