Key result
Reconstructed FFR and iFR using wave separation analysis demonstrated high correlations with conventional FFR (r = 0.992) and iFR (r = 0.930).
Why the study?
The study aimed to predict wedge pressure using the backward wave via wave separation analysis to reflect its effect on FFR and iFR and identify the relationship between the two indices.
Does reconstructed FFR and iFR using wave separation analysis correlate with conventional FFR and iFR in coronary stenosis?
Observational (n=9)
No
Does reconstructed FFR and iFR using wave separation analysis correlate with conventional FFR and iFR in coronary stenosis?
Reconstructed FFR and iFR using wave separation analysis show excellent agreement with conventional indices, suggesting both are inferred from the same physiological formula under different conditions.
Background and objectives Fractional flow reserve (FFR) and instantaneous wave-free ratio (iFR) are the two most commonly used coronary indices of physiological stenosis severity based on pressure. To minimize the effect of wedge pressure (P wedge ), FFR is measured during hyperemia conditions, and iFR is calculated as the ratio of distal and aortic pressures (P d /P a ) in the wave-free period. The goal of this study was to predict P wedge using the backward wave (P back ) through wave separation analysis (WSA) and to reflect the effect of P wedge on FFR and iFR to identify the relationship between the two indices. Methods An in vitro circulation system was constructed to calculate P wedge . The measurements were performed in cases with stenosis percentages of 48, 71, and 88% and with hydrostatic pressures of 10 and 30 mmHg. Then, the correlation between P back by WSA and P wedge was calculated. In vivo coronary flow and pressure were simultaneously measured for 11 vessels in all patients. The FFR and iFR values were reconstructed as the ratios of forward wave at distal and proximal sites during hyperemia and at rest, respectively. Results Based on the in vitro results , the correlation between P back and P wedge was high (r = 0.990, p < 0.0001). In vivo results showed high correlations between FFR and reconstructed FFR (r = 0.992, p < 0.001) and between iFR and reconstructed iFR (r = 0.930, p < 0.001). Conclusions Reconstructed FFR and iFR were in good agreement with conventional FFR and iFR. FFR and iFR can be expressed as the variation of trans-stenotic forward pressure, indicating that the two values are inferred from the same formula under different conditions.
No takes yet. Share an insight, caveat, or question.
Min et al. (2021) conducted an observational in Coronary artery stenosis (n=9). Reconstructed FFR and iFR using wave separation analysis vs. Conventional FFR and iFR was evaluated on Correlation between reconstructed and conventional indices (r = 0.992 (FFR) and r = 0.930 (iFR), p=<0.001). Reconstructed FFR and iFR using wave separation analysis demonstrated high correlations with conventional FFR (r = 0.992) and iFR (r = 0.930).