Answers are grounded in this paper and cite their sources.
This paper illustrates the results of long-term tests on the stability of the output signal of the solid electrolyte amperometry sensor when measuring the hydrogen concentration in the H2 + N2 gaseous mixture. The obtained experimental data verify the stability and reproducibility of the sensor output signal for hydrogen concentration measurements in the nitrogen-containing gaseous mixture during 8000 h of operation. The output signal drift, i.e., the limiting current value, was insignificant, less than ± 5 %. The sensor operation was performed at 3 temperature shifts with different time intervals; these changes did not have any impact either on the sensor integrity or on its operation. The structure of the solid electrolyte sensor, intermediate solid electrolyte / electrode layer and electrodes did not undergo any significant changes during operation. The dynamic characteristics of the sensor, the response time in particular, remained stable during the operation.
No takes yet. Share an insight, caveat, or question.
Kalyakin et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: