Answers are grounded in this paper and cite their sources.
A battery powered, portable laser-induced breakdown spectrometry instrument with a low pulse energy laser and a low resolution, non-gated spectrometer is evaluated. Improvements in the instrument probe design were accomplished through studies of the plasma spatial development, lens-to-sample distance effects and spatial filtering. The prototype was evaluated on paint, steel, ores, and organic samples. Limits of detection ranged from 0.016% for Mn in NIST SRM steel to 0.13% for Ca in NIST SRM organic samples. Acceptable precision (0.4–4.9%) was obtained for steel, ore and organic SRM samples. The poorer precision (4.0–44.1%) obtained for the detection of Pb in paint can be attributed to the heterogeneity of the samples.
No takes yet. Share an insight, caveat, or question.
Castle et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: