Answers are grounded in this paper and cite their sources.
We developed a method to separate and simultaneously quantitate eight phospholipid classes, phosphatidylglycerol (PG), phosphatidylethanolamine (PE), phosphatidylinositol (PI), phosphatidylserine (PS), phosphatidic acid (PA), phosphatidylcholine (PC), sphingomyelin (SM), and lysophosphatidylcholine (LPC). A chloroform/methanol/1‐propanol/ammonium hydroxide/water solvent system was used as the HPLC mobile phase. This condition was set to fulfill a fit in a calibration curve and adequate separation of phospholipids. We evaluated our analytical method for precision, accuracy, and recovery of the phospholipids for human high‐density lipoprotein (HDL). The precision, accuracy, and recovery ranged from 1.4 to 22.6%, 26.9 to +30.4%, and 56.0 to 99.3%, respectively.
No takes yet. Share an insight, caveat, or question.
Hayakawa et al. (2005) studied this question.