Key result
Permanent His bundle pacing achieves ~86% procedural success with stable thresholds at follow-up.
Why the study?
There is limited experience with His bundle pacing at centres initially adopting the procedure, particularly in the Indian setting.
Is permanent His bundle pacing feasible and safe for achieving physiological pacing in routine clinical practice among patients requiring permanent pacing or cardiac resynchronization therapy?
Observational (n=22)
No
Is permanent His bundle pacing feasible and safe for achieving physiological pacing in routine clinical practice among patients requiring permanent pacing or cardiac resynchronization therapy?
This first reported experience from India demonstrates that permanent His bundle pacing is a feasible and safe physiological pacing option with an 86% implant success rate, stable capture thresholds, and LVEF improvement in patients with LV dysfunction.
Supports feasibility in this small series; hypothesis-generating and requires randomized confirmation before wider adoption.
There is a paucity of experience regarding His bundle pacing (HBP) at laboratories initially attempting the procedure, especially in the Indian scenario. Patient who underwent HBP were selected for pacing therapy or in lieu of cardiac resynchronization therapy (CRT) at a single center. Among 22 patients attempted, 19 patients underwent successful implant, achieving selective HBP in 14 patients. There was a significant improvement in left ventricular ejection fraction (LVEF) (49.3 ± 9.3 vs. 36.7 ± 9.2) in the LV dysfunction subgroup (n = 6). Over a follow-up of 15 ± 6.5 months, thresholds were stable in all except one patient, and there was no requirement of lead revision. In summary, we found that HBP is a feasible option for achieving physiological pacing.
No takes yet. Share an insight, caveat, or question.
Sarkar et al. (2019) conducted an observational in Conduction disease requiring permanent pacing or indication for cardiac resynchronization therapy (CRT) (n=22). Permanent His bundle pacing (HBP) using Select Secure 3830 lead and Select Site C315 sheath was evaluated on Procedural success (defined as a pacing threshold for His capture of ≤2.5 V at 1 ms).