Key result
Sirolimus-coated stents yield minimal neointimal hyperplasia with 0% restenosis and major events at 8 months.
Answers are grounded in this paper and cite their sources.
Demonstrates minimal neointimal proliferation in patients with angina, indicating that sirolimus-coated stents effectively inhibit in-stent restenosis.
BACKGROUND: Restenosis remains an important limitation of interventional cardiology. Therefore, we aimed to determine the safety and efficacy of sirolimus (a cell-cycle inhibitor)-coated BX Velocity stents. METHODS AND RESULTS: Thirty patients with angina pectoris were electively treated with 2 different formulations of sirolimus-coated stents (slow release [SR], n=15, and fast release [FR], n=15). All stents were successfully delivered, and patients were discharged without clinical complications. Independent core laboratories analyzed angiographic and 3D volumetric intravascular ultrasound data (immediately after procedure and at 4-month follow-up). Eight-month clinical follow-up was obtained for all patients. There was minimal neointimal hyperplasia in both groups (11.0+/-3.0% in the SR group and 10.4+/-3.0% in the FR group, P:=NS) by ultrasound and quantitative coronary angiography (in-stent late loss, 0.09+/-0.3 mm [SR] and -0.02+/-0.3 mm [FR]; in-lesion late loss, 0.16+/-0.3 mm [SR] and -0.1+/-0.3 mm [FR]). No in-stent or edge restenosis (diameter stenosis >or=50%) was observed. No major clinical events (stent thrombosis, repeat revascularization, myocardial infarction, or death) had occurred by 8 months. CONCLUSIONS: The implantation of sirolimus-coated BX Velocity stents is feasible and safe and elicits minimal neointimal proliferation. Additional placebo-controlled trials are required to confirm these promising results.
No takes yet. Share an insight, caveat, or question.
Sousa et al. (2001) studied Angina pectoris / Coronary artery disease (n=30). Sirolimus-coated BX Velocity stents (slow release [SR]) vs. Sirolimus-coated BX Velocity stents (fast release [FR]) was evaluated on Neointimal hyperplasia (% by intravascular ultrasound at 4 months) (p=NS). Sirolimus-coated BX Velocity stents elicited minimal neointimal hyperplasia at 4 months (11.0% slow release vs 10.4% fast release, P=NS) with 0% restenosis or major events at 8 months.
Synapse has enriched 10 related papers. Consider them for comparative context: