Key result
A 77 GHz FMCW mm-wave radar system combined with an artificial neural network achieved a mean test accuracy of 75% for predicting the presence of arrhythmia.
Why the study?
Does a 77 GHz FMCW mm-wave radar combined with an artificial neural network accurately measure vital signs and detect arrhythmias?
Cross-Sectional (n=15)
Does a 77 GHz FMCW mm-wave radar combined with an artificial neural network accurately measure vital signs and detect arrhythmias?
A novel 77 GHz FMCW mm-wave radar system combined with an artificial neural network can detect arrhythmias with 75% accuracy in a small test cohort, offering a potential non-contact monitoring solution.
May enable contactless arrhythmia screening; leaves open prospective clinical validation before adoption.
A non-contact, non-invasive monitoring system to measure and estimate the heart and breathing rate of humans using a frequency-modulated continuous wave (FMCW) mm-wave radar at 77 GHz is presented. A novel diagnostic system is proposed which extracts heartbeat phase signals from the FMCW radar (reconstructed using Fourier series analysis) to test a three-layer artificial neural network model to predict the presence of arrhythmia in individuals. The effect of person orientation, distance of measurement and movement was analyzed with respect to a reference device based on statistical measures that include number of outliers, mean, mean squared error (MSE), mean absolute error (MAE), median absolute error (medAE), skewness, standard deviation (SD) and R-squared values. The individual oriented in front of the radar outperformed almost all other orientations for most distances with an expected d = 90 cm and d = 120 cm. Furthermore, it was found that the heart rate that was measured while walking and the breathing rate which was measured for a motionless individual generated results with the lowest SD and MSE. An artificial neural network (ANN) was trained using the MIT-BIH database with a training accuracy of 93.9 % and an R2 value = 0.876. The diagnostic tool was tested on 15 subjects and achieved a mean test accuracy of 75%.
No takes yet. Share an insight, caveat, or question.
Iyer et al. (2022) conducted a cross-sectional in Arrhythmia (n=15). FMCW mm-wave radar at 77 GHz with artificial neural network vs. Reference device was evaluated on Mean test accuracy for arrhythmia prediction. A 77 GHz FMCW mm-wave radar system combined with an artificial neural network achieved a mean test accuracy of 75% for predicting the presence of arrhythmia.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: