A noninvasive method using an inversion recovery turbo-FLASH for dynamic measurement of the arterial input function represented by the bolus passage of Gd-DTPA in the descending aorta is presented, and the results are compared with the input function obtained by arterial blood samples. A good accordance between the two input functions was found, indicating that it is possible to measure the input function to the myocardium using MRI. A variation between the two concentration curves of 5% at upslope, 2.7% at peak point, and < 7% at downslope was found. The study also indicates that a short inversion time < 250 ms has to be used to ensure correct measurement of peak concentration.
No takes yet. Share an insight, caveat, or question.
Fritz‐Hansen et al. (1996) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: