Specific combinations of ICD sensing algorithms and integrated bipolar leads can cause inappropriate shocks during atrial flutter with intermittent pacing.
Answers are grounded in this paper and cite their sources.
May trigger inappropriate shocks with this ICD-lead pair in atrial flutter; case report leaves open broader applicability.
INTRODUCTION: Inappropriate discharges of an implantable cardioverter defibrillator (ICD) are troublesome to the patient and sometimes a difficult task for the physician trying to identify and treat the cause. METHODS AND RESULTS: For the first time, we report a mechanism of inappropriate ICD discharges during episodes of atrial flutter with a slow ventricular response and intermittent antibradycardia pacing. The episodes occurred in two patients and were triggered by the unique sensing algorithm of the Ventritex Cadence V-100 in combination with the tripolar CPI Endotak 072 transvenous defibrillation lead, which provides integrated bipolar sensing. CONCLUSION: Besides treatment of the underlying arrhythmia, reprogramming of the device, an electrode position far away from the atria, and true bipolar sensing will enhance the performance of ICD systems with respect to the episodes described here. In addition, more flexible sensing algorithms may, in the future, prevent this overall rare complication.