Key result
Non-standard ECG electrode placements yield recognizable waveforms with ~27% median R-wave sensitivity.
Why the study?
Expectations for wearable cardiovascular monitoring devices to be unobtrusive and miniaturized require developing smarter devices and better methods for ECG signal acquisition and processing.
Do non-standard thoracic and single-arm electrode placement configurations provide acceptable ECG signal quality for joint ECG and impedance cardiography monitoring in a wearable device context?
Do non-standard thoracic and single-arm electrode placement configurations provide acceptable ECG signal quality for joint ECG and impedance cardiography monitoring in a wearable device context?
Answers are grounded in this paper and cite their sources.
This proof-of-concept study demonstrates the feasibility of acquiring recognizable ECG waveforms from non-standard thoracic and single-arm electrode locations shared with impedance cardiography sensors for compact wearable device design.
BACKGROUND: Wearable technologies for monitoring cardiovascular parameters, including electrocardiography (ECG) and impedance cardiography (ICG), propose a challenging research subject. The expectancy for wearable devices to be unobtrusive and miniaturized sets a goal to develop smarter devices and better methods for signal acquisition, processing, and decision-making. METHODS: In this work, non-standard electrode placement configurations (EPC) on the thoracic area and single arm were experimented for ECG signal acquisition. The locations were selected for joint acquisition of ECG and ICG, targeted to suitability for integrating into wearable devices. The methodology for comparing the detected signals of ECG was developed, presented, and applied to determine the R, S, and T waves and RR interval. An algorithm was proposed to distinguish the R waves in the case of large T waves. RESULTS: Results show the feasibility of using non-standard EPCs, manifesting in recognizable signal waveforms with reasonable quality for post-processing. A considerably lower median sensitivity of R wave was verified (27.3%) compared with T wave (49%) and S wave (44.9%) throughout the used data. The proposed algorithm for distinguishing R wave from large T wave shows satisfactory results. CONCLUSIONS: The most suitable non-standard locations for ECG monitoring in conjunction with ICG were determined and proposed.
No takes yet. Share an insight, caveat, or question.
Metshein et al. (2022) studied Healthy volunteer for cardiovascular signal monitoring (n=1). Non-standard electrode placement configurations (7 thoracic and 1 single-arm) for joint ECG and ICG signal acquisition vs. Standard 3-electrode ECG system on the thorax was evaluated on ECG signal quality and median sensitivity of R, S, and T wave detection. Non-standard electrode placement configurations yielded recognizable ECG waveforms with a median R-wave sensitivity of 27.3% compared with 49% for T-waves and 44.9% for S-waves.