Key result
The proposed MTV and MWTV approaches demonstrated better denoising performance for EEG motion artifacts with average ASNR values of 29.12 dB and 29.29 dB, respectively, compared to existing techniques.
Why the study?
Filtering out noise and artifacts from EEG signals is vital for the automated detection of neurological disorders.
Do MTV and MWTV filtering schemes improve the removal of motion artifacts from single channel EEG signals compared to existing techniques?
Do MTV and MWTV filtering schemes improve the removal of motion artifacts from single channel EEG signals compared to existing techniques?
MTV and MWTV filtering schemes effectively remove motion artifacts from single-channel EEG signals, improving signal quality for automated detection of neurological disorders.
The electroencephalogram (EEG) signal is contaminated with various noises or artifacts during recording. For the automated detection of neurological disorders, it is a vital task to filter out these artifacts from the EEG signal. In this paper, we propose two novel approaches for the removal of motion artifact from the single channel EEG signal. These methods are based on the multiresolution total variation (MTV) and multiresolution weighted total variation (MWTV) filtering schemes. The multiresolution analysis using the discrete wavelet transform (DWT) helps to segregate the EEG signal into various subband signals. The total variation (TV) and weighted TV (WTV) are applied to the approximation subband signal. The filtered approximation subband signal is evaluated based on the difference between the noisy approximation subband signal and the output of the TV or WTV filter. The processed EEG signal is obtained using the multiresolution wavelet-based reconstruction. The difference in the signal to noise ratio (ASNR) and the percentage of reduction in correlation coefficients (η) is used for evaluating the diagnostic quality of the processed EEG signal. The experimental results demonstrate that the proposed MTV and MWTV approaches have better denoising performance with (average ASNR, and average η) values of (29.12 dB and 68.56%) and (29.29 dB and 67.51%), respectively, as compared to the existing techniques.
No takes yet. Share an insight, caveat, or question.
Gajbhiye et al. (2019) studied EEG signal artifacts. Multiresolution total variation (MTV) and multiresolution weighted total variation (MWTV) filtering schemes vs. existing techniques was evaluated on Difference in the signal to noise ratio (ASNR) and the percentage of reduction in correlation coefficients (η). The proposed MTV and MWTV approaches demonstrated better denoising performance for EEG motion artifacts with average ASNR values of 29.12 dB and 29.29 dB, respectively, compared to existing techniques.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: