Why the study?
What is the minimal pneumatic tourniquet pressure necessary to produce a bloodless field during limb surgery?
What is the minimal pneumatic tourniquet pressure necessary to produce a bloodless field during limb surgery?
Minimal tourniquet pressures of 200 mm Hg for upper limbs and 250 mm Hg for lower limbs are adequate for average-sized normotensive patients, potentially reducing complications from excessive pressure.
Answers are grounded in this paper and cite their sources.
May support lower tourniquet pressures in average patients; cross-sectional data leaves open need for randomized validation.
Complications attributable to direct pressure may result from the use of pneumatic tourniquets during surgical procedures. Traditional estimates have determined the pressures employed rather than the minimal pressure necessary to produce a bloodless field. To determine this pressure, pre-operative and post-operative systolic blood pressures and the tourniquet pressure at which capillary bleeding occurred were measured in a group of patients undergoing elective surgery of the upper and lower limbs. From these results two equations were derived, one for each of the upper and lower limbs, which give the minimum tourniquet pressures to produce bloodless fields. In an average sized, normotensive patient, 200mm Hg was found to be adequate for the upper limb and 250mm Hg for the lower limb.