Answers are grounded in this paper and cite their sources.
The critical design aim for a sodium/proton coil is to maximize sodium sensitivity and transmit field homogeneity while simultaneously providing adequate proton sensitivity and homogeneity. While most dual-frequency coils use lossy high-impedance trap circuits or PIN diodes to allow dual-resonance, we explored a nested-coil design for sodium/proton knee imaging at 7 T. A stand-alone eight-channel sodium receive array was implemented without standard dual-resonance circuitry to provide improved sodium signal-to-noise ratio. A detunable sodium birdcage was added for homogeneous sodium excitation and a four-channel proton transmit-receive array was added to provide anatomical reference imaging and B0 shimming capabilities. Both additional modules were implemented with minimal disturbance to the eight-channel sodium array by managing their respective resonances and geometrical arrangement. In vivo sodium signal-to-noise ratio was 1.2-1.7 times greater in the developed eight-channel array than in a mononuclear sodium birdcage coil, whereas the developed four-channel proton array provided signal-to-noise ratio similar to that of a commercial mononuclear proton birdcage coil.
No takes yet. Share an insight, caveat, or question.
Brown et al. (2012) studied this question.