Answers are grounded in this paper and cite their sources.
This survey focused on clinical reports of polyethylene wear and osteolysis in total hip replacements. With regard to documentation of clinical wear-rates, 57 publications were reduced to an analysis of 8 reports of THR series, including the incidence of osteolysis. A direct correlation was found among volumetric wear-rates, incidence of osteolysis and revision rates in THR concepts of the 1983-1987 era. As volumetric wear rate increased, the incidence of osteolysis and revision rates increased. With regard to our grading system for volumetric wear, with follow-up in the 4-15 year range, osteolysis was rare in group A (wear = 0-80 mm3/year), ranging from 6% to 31% in group B (wear 40-80 mm3/year) and from 21% to 100% in group C (wear > 140 mm3/year). With regard to cup design, the optimal low-wear group had mainly cemented polyethylene cups with 22 and 28 mm head sizes. The mid-wear group B had metal-backed cemented and uncemented cups, with 28 mm head size, and the high-risk group C had only uncemented, metal-backed cups, with the highest wear in the 32 mm head size. Less than 10 years of follow-up did not distinguish adequately between different designs of THR, except in a few cases which had early failures due to material or design deficiencies. Overall, the cemented all-polyethylene cup combined with the smaller ball head proved to be better.
No takes yet. Share an insight, caveat, or question.
Oparaugo et al. (2001) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: