Why the study?
To prospectively assess the presence and degree of tricuspid regurgitation after right ventricular lead placement for a permanent pacemaker or implantable cardioverter defibrillator.
Does right ventricular endocardial lead placement cause tricuspid regurgitation in patients undergoing pacemaker or ICD implantation?
Does right ventricular endocardial lead placement cause tricuspid regurgitation in patients undergoing pacemaker or ICD implantation?
Endocardial lead placement for pacemakers or ICDs is associated with the development of tricuspid regurgitation in 10% of patients at 6 months.
TR after RV lead placement may warrant monitoring; leaves open causal impact on device selection or outcomes.
Objective: To assess prospectively the presence and degree of tricuspid regurgitation (TR) after right ventricular (RV) lead placement utilizing 2-dimensional and doppler echocardiography in a group of patients implanted with Permanent pace-maker or implantable cardioverter defibrillator (ICD). Study Design: Descriptive cross sectional study. Place and Duration of Study: Study was conducted at OPD of AFIC/NIHD Rawalpindi from Jan to Jun 2019. Methodology: One hundred patients after cardiac stimulation system implantation were included in this study. Patients with severe valve disease, heart failure, congenital heart disease, pre-existing pulmonary hypertension and presence of moderate or severe tricuspid regurgitation were excluded. M-mode, 2 –dimensional and doppler echocardiographic studies were performed. Continuous-wave doppler measurements were made from apical four-chamber view in order to obtain maximum tricuspid flow velocities. A complete echocardiographic study was performed following device implantation. Echocardiographic measurements were repeated at 6-months of follow-up period. Those patients who developed tricuspid regurgitation were further assessed by color doppler, taking into account density and contour of the jet on continuous-wave doppler. Tricuspid regurgitation severity was classified into three groups: mild, moderate and severe according to the recommendations for non-invasive evaluation of native valvular regurgitation by the American Society of Echocardiography. The collected data was analysed by using SPSS-23. Results: Out of 100 patients, 67 (67%) were males and 33(33%) were females with the mean age of 65.29 ± 12.02 years. All the patients had normal chambers dimensions before the procedure. 65 (65%) had leads screwed to RV apex while 35 (35%) had through RV septum. 83 (83%) patients received a dual chamber device, while 17 (17%) patients got single chamber (ventricular) device implanted. Following device implantation after 6 months mild TR was noted in 4 (4%), moderate TR in 4 (4%) and severe TR 2 (2%) in cases. Conclusion: Tricuspid incompetence following endocardial leads implantation is not rare and occurred in approximately 10% of our patients. This complication may be preventable, because it is likely due to the interference of the endocardial lead with the TV.
No takes yet. Share an insight, caveat, or question.
Asad et al. (2021) studied this question.