Why the study?
Does a machine learning strategy using PTT parameters improve the accuracy of noninvasive cuffless blood pressure measurement?
Does a machine learning strategy using PTT parameters improve the accuracy of noninvasive cuffless blood pressure measurement?
A machine learning approach using pulse transit time from ECG and PPG signals can reduce errors in noninvasive cuffless blood pressure measurement in computer simulations.
ML cuffless BP via ECG/PPG feasible in simulations; leaves open prospective clinical validation.
A technique based on a machine learning approach is suggested and studied for blood pressure measurements. The proposed technique uses a noninvasive cuffless approach for blood pressure evaluation. In order to extract blood pressure data using this noninvasive cuffless method, pulse wave velocity or pulse wave travel time (PTT) are estimated by both signal processing of electrocardiogram (ECG) and photoplethysmogram (PPG) data records. For study performed by computer simulations, the ECG and PPG records were taken from an open database. Errors arising both for systolic and diastolic arterial pressure evaluation were estimated. Computer simulation results indicate that using machine learning strategy and using only PTT parameters provide a considerable decrease in root mean square errors both for systolic and diastolic human blood pressure data.
No takes yet. Share an insight, caveat, or question.
Viunytskyi et al. (2022) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: