Answers are grounded in this paper and cite their sources.
Objective Identify opportunities to reduce adverse events of magnetic resonance imaging (MRI) with hearing implants that require head wraps. Methods Retrospective review of MRIs ordered for patients with MR Conditional devices between June 2022 and October 2024 at a tertiary center. Results Thirty-one patients (27 cochlear and four Osia implants) had 38 MRIs ordered. Nine orders (24%) were canceled because discussion clarified that imaging was not needed (two orders, 5%), could be substituted by CT imaging (six orders, 16%), or patient declined (one order, 3%). Twenty-nine (76%) orders resulted in scans performed with head wraps, 5 (17%) of which resulted in magnet site pain without long term complication and early termination of four scans (14%). Conclusion A quarter of MRI orders were avoidable through pre-imaging review, and among scans performed, magnet site pain and early termination were not infrequent. The mnemonic “MRI” offers a practical framework to reduce these adverse events: Minimize unnecessary MR imaging through multidisciplinary pre-authorization, Reprioritize critical sequences early in case of early termination, and Immobilize the internal magnet to reduce pain and device displacement.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2026) studied this question.