Detecting interfacial slippage from weak nonlinear vibration signatures presents a fundamental challenge in contact interface measurement. Under conventional vibration excitation alone, interfaces with partial contact degradation exhibit near-linear dynamic behaviour, and higher-order harmonic components cannot be reliably identified. This study presents a preload-assisted harmonic measurement methodology in which axial compressive load is applied as a measurement conditioning variable during torsional harmonic excitation, driving the interface toward its slippage threshold and amplifying the nonlinear harmonic response. The methodology was validated using press-fit bone-implant models with varied cortical thickness, providing a controlled range of interfacial contact conditions representative of progressive contact degradation. Push-in tests confirmed that cortical thinning reduces contact pressure and load-bearing capacity, lowering the threshold for interfacial slippage. Vibration measurements at the first torsional resonance demonstrated that axial preload significantly amplifies higher-order harmonic components, with harmonic amplitude serving as a direct indicator of interfacial slip events. The proposed measurement protocol provides a novel approach to contact interface assessment, applicable beyond the biomedical context to any system in which interfacial degradation must be identified from weak nonlinear signatures under conventional excitation conditions.