Key result
Anticoagulant counter ion changes show no impact on bicalutamide LC-MS/MS bioanalytical assay stability.
Answers are grounded in this paper and cite their sources.
BACKGROUND: In the past several years, the impact of changing counter ions while keeping the same anticoagulant in bioanalytical LC-MS/MS methods has become a highly discussed topic. In order to confirm that there is no impact from counter ions, matrix effect and stability evaluations were performed on bicalutamide LC-MS/MS bioanalytical methods. RESULTS: Independently from the anticoagulant counter ion used, the matrix effect evaluation met acceptance criteria, even when using conditions expected to increase matrix effect, such as protein precipitation with an analog internal standard. Freeze-thaw along with storage stabilities, namely short- and long-term, demonstrated less than 8% deviation regardless of the counter ion used. CONCLUSION: Differences in the anticoagulant counter ion used has no impact on the bicalutamide bioanalytical LC-MS/MS method.
No takes yet. Share an insight, caveat, or question.
Bergeron et al. (2012) studied this question. Anticoagulant counter ions in bicalutamide LC-MS/MS bioanalytical methods vs. Different anticoagulant counter ions was evaluated on Matrix effect and freeze-thaw/storage stability in bicalutamide LC-MS/MS bioanalytical methods. Changing the anticoagulant counter ion had no impact on the bicalutamide LC-MS/MS bioanalytical method, showing <8% deviation in freeze-thaw and storage stability.