Key result
Inadequate sampling tubing size and flow rates distort in vivo indicator dilution curves.
Key points are not available for this paper at this time.
Answers are grounded in this paper and cite their sources.
This paper describes the distorting effects produced by various sizes of polyethylene tubing and various flow rates on a square wave input of dye concentration. The importance of an accurate sampling system when recording dilution curves for analysis of their shape is emphasized and the distortion caused by inadequate sampling systems is demonstrated in vivo.
No takes yet. Share an insight, caveat, or question.
Lacy et al. (1957) studied this question. Inadequate sampling systems with varied polyethylene tubing sizes and flow rates distort the shape of indicator dilution curves in vivo.
Synapse has enriched 10 related papers. Consider them for comparative context: