Purpose: This biomechanical study evaluated the effect of intramedullary screw diameter and length relative to 3-point bending force and torsional force when used to stabilize metacarpal shaft fractures. Methods: Transverse osteotomies were made in the proximal metacarpal shaft in 36 middle finger metacarpal fourth-generation composite Sawbones. To compare screw diameters, antegrade intramedullary screws of 30-mm length were placed in 6 metacarpals, which included 4.7-mm Acutrak 2, Standard Acutrak 2 (4.0 mm), and Mini-Acutrak 2 (3.5 mm) screws. To compare screw lengths, metacarpals were fixated with Standard Acutrak 2 screws of 26, 30, or 34 mm in length, with screw tips bypassing the osteotomy by 6, 10, or 14 mm, respectively. A 6 degrees of freedom robot was used for torsional and 3-point bending testing. Conclusions: The results demonstrated significantly higher torsional strength and resistance to 3-point bending with larger intramedullary screw diameters. Further, when selecting the intramedullary screw length, the screw tip should pass at least 10 mm beyond the fracture. Clinical Relevance: This study provided biomechanical evidence to guide surgeons in selecting intramedullary screw diameter and length for treating metacarpal fractures.