A 70-year-old-woman was admitted to the ICU with septic shock and was treated with fluid resuscitation, broad-spectrum antibiotics, and initiation of vasopressors. An attempt was made to insert a central venous catheter (CVC) in the right subclavian vein. The patient was placed in the Trendelenburg position, with head turned contralaterally and neutral shoulder position. Venous access was obtained with the introducer needle, using landmarks alone without complication, and the guidewire inserted to a depth that was later determined to be approximately 35 cm. The finder needle was withdrawn, and the tract dilated without difficulty. The catheter was smoothly inserted over the guidewire without resistance to a depth of approximately 17 cm. However, significant resistance was encountered in attempting to withdraw the guidewire through the catheter, and the procedure was aborted. Multiple portable bedside radiographs were obtained (Figures 1A and 1B) and showed the guidewire turning cephalad up the internal jugular vein, into the skull via the sigmoid dural sinus, traveling across the right transverse sinus into the left transverse sinus (Figure 1C). The neurosurgical service was consulted, and the wire was removed using significant manual force. After removal, a computed tomography venogram revealed no vascular complications.
No takes yet. Share an insight, caveat, or question.
Trimble et al. (2018) studied this question.