Why the study?
Does a stereoscopic DSA system accurately measure absolute blood flow rates in a vessel phantom compared to volumetric or electromagnetic flowmeter calibration?
Does a stereoscopic DSA system accurately measure absolute blood flow rates in a vessel phantom compared to volumetric or electromagnetic flowmeter calibration?
A stereoscopic DSA system demonstrates reasonable accuracy in measuring absolute blood flow rates in a vessel phantom, particularly for continuous flow.
Phantom validation supports DSA flowmetry development; leaves open clinical translation pending in vivo pulsatile testing.
We used a stereoscopic digital subtraction angiography (DSA) system to measure absolute blood flow rates in vessels. The magnification factor and the three-dimensional orientation of a selected vessel are obtained from automated analysis of stereoscopic DSA images. The cross-sectional area of the vessel is determined from the vessel diameter, which is measured with an iterative deconvolution technique. The time required for fluid to flow through a selected segment of a vessel is determined from the automated analysis of contrast medium 'time-density' curves. The effectiveness of these combined techniques was demonstrated in measurement of rates of both continuous and pulsatile flow in a vessel phantom, with the actual flow rate calibrated volumetrically or by an electromagnetic flowmeter. We have obtained accuracies in measured flow rates of approximately 5% and 18% for continuous and pulsatile flow respectively.
No takes yet. Share an insight, caveat, or question.
Fencil et al. (1989) studied this question.