INTRODUCTION: This study was designed to evaluate short-term physiologic outcomes of transitioning neonates with bronchopulmonary dysplasia (BPD) from intensive care unit (ICU) ventilators to both the Trilogy 202 (Philips Healthcare, Andover, MA) and LTV 1200 (CareFusion, Yorba Linda, CA) subacute ventilators. METHODS: ), ventilator response times, and the percentage of spontaneous breaths that triggered the ventilator at 20 min in each subject while being supported with each of the different subacute ventilators. All six infants transitioned successfully from the ICU ventilator to the Trilogy 202 ventilator. CONCLUSION: In this small group of infants with BPD, the Trilogy 202 ventilator performed better than the LTV 1200. The improved subject efforts, per cent subject triggering, and response times observed with the Trilogy are likely related to differences in triggering algorithms, location of triggering mechanisms, and gas delivery system performance within the ventilators. These pilot data may be useful for informing future clinical study design and understanding differences in the level of support provided by different subacute ventilators in infants with BPD.