Signal intensity in magnetic resonance imaging is described by the solution to the Bloch Equations. Image behavior has been calculated for the variable angle gradient refocusing and conventional spin echo techniques. It was found that by proper manipulation of the excitation angle (and echo time in spin echo sequences) that the imaging time can be reduced up to twenty-fold for the gradient refocusing technique and four-fold for conventional spin echo imaging. The contrast obtained is comparable to that acquired in conventional MRI. Imaging experiments performed on phantoms and human heads are in good agreement with theoretical results.
No takes yet. Share an insight, caveat, or question.
Vinitski et al. (1987) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: