Answers are grounded in this paper and cite their sources.
PURPOSE: To compare flap thickness reproducibility of the femtosecond laser and two mechanical microkeratomes. METHODS: Flap thickness for all eyes was measured as the difference between the preoperative (day of surgery) full corneal thickness and post-flap creation central stromal bed thickness using ultrasonic pachymetry. Flap thickness values produced by three different microkeratome systems were compared for accuracy and reproducibility. RESULTS: For 99 flaps created using the IntraLase FS laser with an intended thickness of 110 microm, the mean achieved thickness was 119 +/- 12 microm (range: 82 to 149 microm). In 100 eyes treated with the Moria LSK-1 microkeratome with an intended flap thickness of 160 microm, the mean achieved thickness was 130 +/- 19 microm (range: 71 to 186 microm). In 135 eyes treated with the Moria M2 microkeratome with an intended flap thickness of 130 microm, mean thickness was 142 +/- 24 microm (range: 84 to 203 microm). The standard deviation and range of corneal flap thickness created with the IntraLase FS laser was significantly smaller than either mechanical microkeratome (P < .0001). CONCLUSIONS: When compared to two commonly used mechanical microkeratomes, mean achieved flap thickness was more reproducible with the IntraLase FS laser, reducing the comparative risk of overly thick flaps.
No takes yet. Share an insight, caveat, or question.
Talamo et al. (2006) studied this question.