Why the study?
To evaluate the mid-term outcomes of the Pulsta transcatheter pulmonary valve for percutaneous pulmonary valve implantation in native right ventricular outflow tract lesions.
Does percutaneous pulmonary valve implantation with the Pulsta valve improve hemodynamics and RV volume in patients with severe pulmonary regurgitation in the native RVOT?
Does percutaneous pulmonary valve implantation with the Pulsta valve improve hemodynamics and RV volume in patients with severe pulmonary regurgitation in the native RVOT?
The Pulsta transcatheter pulmonary valve demonstrates good mid-term hemodynamic effectiveness and safety for treating severe pulmonary regurgitation in the native RVOT.
Answers are grounded in this paper and cite their sources.
Supports Pulsta valve feasibility in native RVOT; leaves open durability and comparative efficacy questions.
Objectives The aim of this study is to present the mid‐term outcomes of Pulsta valve. Background The Pulsta valve is a Self‐expandable knitted nitinol‐wire stent mounted with a treated tri‐leaflet α‐Gal‐free porcine pericardial valve for percutaneous pulmonary valve implantation (PPVI) in patients with native right ventricular outflow tract (RVOT) lesions. Methods A multi‐center clinical trial using Pulsta valve® was designed for patients with severe pulmonary regurgitation (PR) in the native RVOT in multiple centers in South Korea and 25 patients were enrolled. Before PPVI, severe PR (mean PR fraction: 45.5 ± 6.9%) and enlarged RV volume (mean indexed RV end‐diastolic volume; 169.7 ± 13.0 ml/m 2 ) was present. The mean age was 21.6 ± 6.6 years old. Results All patients were successfully implanted with 26, 28, or 32 mm diameter of Pulsta valve loaded on the 18 or 20 French delivery catheters. At 6 months follow up, indexed RV end‐diastolic volume was decreased to 126.9 ± 16.9 ml/m 2 . At mean 33.1 ± 14.3 months follow‐up, the mean value of mean pressure gradient in Pulsta valve was 6.5 ± 3.0 mmhg without significant PR. There was no serious device‐related adverse event. Conclusions A multi‐center clinical trial was completed successfully with planned Pulsta valve implantation and demonstrated good mid‐term effectiveness without device‐related serious adverse events.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: