We report on the design, fabrication, and test of a passive pressure transensor. The sensor uses the self-resonant frequency modulation of an integrated coil to detect the pressure variations. This modulation is generated by the relative displacement of a ferrite core attached to a silicone rubber membrane. This scheme simplifies the packaging of the passive transensor by removing the requirement for a separate capacitor to form the resonator. A 30-turn 1.7-μH coil having dimensions of 3×3 mm2is used in a prototype design yielding a sensitivity of 9.6 kHz/kPa with a cylindrical ferrite core of 0.95-mm diameter and 0.5-mm height. We also present a theoretical model of the sensor that shows good agreement with the experimental data. This model can be a useful tool for further optimization of the senors.
No takes yet. Share an insight, caveat, or question.
Baldi et al. (2003) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: