Compares k-space acquisition strategies in nonuniform MRI, suggesting improvements in image fidelity and efficiency.
PURPOSE: inhomogeneity, and RF coil bandwidth constraints. METHODS: Three acquisition strategies were compared: single-shot turbo spin echo (SS-TSE), multi-echo spin echo (MESE) with repeated phase encoding, and spin echo single-point imaging (SE-SPI) without applied readout (RO) gradients. Experiments were performed on an open, low-field, field-cycling MRI system using a nonuniform electromagnet. Imaging performance was assessed using a phantom and biological samples under matched total scan-time conditions. RF coils with wide- and narrow-bandwidth designs were evaluated to study sensitivity trade-offs. Retrospective k-space undersampling and compressed sensing reconstruction were applied to investigate potential scan-time reduction for all acquisition strategies. RESULTS: SE-SPI consistently achieved larger full width at half maximum (FWHM) values than SS-TSE and MESE, indicating improved edge definition. Edge sharpness was higher along the RO (PE1) direction but comparable along the PE (PE2) direction. SS-TSE provided the fastest k-space acquisition and the highest central k-space SNR but exhibited increased blurring and bandwidth-related signal loss. MESE improved edge definition over SS-TSE while requiring additional polarization cycles. SE-SPI also showed greater robustness to compressed sensing undersampling. CONCLUSION: field-cycling systems by improving image fidelity, maintaining uniform signal intensity, enabling efficient use of high-Q RF coils, and supporting compressed sensing acceleration.
No takes yet. Share an insight, caveat, or question.
Ha et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: