Together with droplet creation, fusion and sorting, the incubation of droplets is one of the most important and essential operations for droplet-based microfluidic assays. This manuscript concerns the development of delay-lines, which are necessary to allow incubation of reactions for precise time periods. We analyze the problems associated with creating delay-lines for incubation in the minute to hour time range, which arise from back-pressure and from the dispersion in the incubation time due to the unequal speeds with which droplets pass through the delay-line. We describe delay-line systems which resolve these problems and demonstrate their use to measure reaction kinetics over several minutes in droplets.
No takes yet. Share an insight, caveat, or question.
Frenz et al. (2008) studied this question.