Key result
FEops HEARTguide underreports aortic root dimensions vs. GE software, altering TAVI sizing in ~23% of cases.
Why the study?
Preprocedural virtual planning of TAVI using FEops HEARTguide may reduce major complications, but clinical validation against standard software and actual procedural outcomes was needed.
Does FEops HEARTguide software accurately predict aortic root dimensions, post-TAVI paravalvular AR, and new conduction abnormalities compared to GE software and clinical outcomes in TAVI patients?
Observational (n=106)
Does FEops HEARTguide software accurately predict aortic root dimensions, post-TAVI paravalvular AR, and new conduction abnormalities compared to GE software and clinical outcomes in TAVI patients?
FEops HEARTguide software predicts post-TAVI AR and conduction abnormalities at a population level but underestimates aortic root dimensions compared to GE software, leading to potential prosthesis sizing differences.
Caution against relying on FEops HEARTguide for individual TAVI sizing; leaves open its population-level predictive value pending prospective validation.
Background Preprocedural virtual planning of transcatheter aortic valve implantation (TAVI) using FEops HEARTguide TM may reduce major TAVI complications. Aims (1) To compare morphometric aortic root measurements by FEops HEARTguide TM and GE software; (2) to validate the accuracy of FEops HEARTguide TM in predicting post‐TAVI paravalvular aortic regurgitation (AR) and new conduction abnormalities. Methods In 106 consecutive TAVI patients, aortic root dimensions essential for prosthesis sizing were measured using both GE software and FEops HEARTguide TM . FEops HEARTguide TM predictions were compared with the severity of post‐TAVI AR assessed by echocardiography, angiography, invasive hemodynamics, and the development of conduction abnormalities assessed by serial electrocardiography. Results Bland‐Altman analysis revealed smaller annulus perimeter, area and coronary artery ostia heights with FEops HEARTguide TM compared with GE software, leading to differences in prosthesis sizing in 23% of cases. Predicted AR was significantly higher in patients with moderate versus none/trace AR on echocardiography (14.3 vs. 2.8 mL/s, p = 0.006) and angiography (23.6 vs. 4.0 mL/s, p = 0.009), though overlap between adjacent AR grades was observed. Predicted AR did not significantly differ between patients with AR index < 25 and ≥ 25 (10.2 vs. 5.7 mL/s, p = 0.079). Contact pressure index was significantly higher in patients with new conduction abnormalities (17.5 vs. 10.0%, p = 0.006), despite overlap at patient level. Conclusions FEops HEARTguide TM generally underestimates aortic root dimensions compared with GE software, affecting prosthesis sizing in a substantial proportion of patients. While predictive at a population level, overlap in parameters predicted by FEops HEARTguide TM is encountered at an individual patient level.
No takes yet. Share an insight, caveat, or question.
Claes et al. (2025) conducted an observational in Transcatheter aortic valve implantation (TAVI) (n=106). FEops HEARTguide vs. GE software was evaluated on Morphometric aortic root measurements and prediction of post-TAVI paravalvular aortic regurgitation and new conduction abnormalities. FEops HEARTguide underestimated aortic root dimensions compared with GE software, altering prosthesis sizing in 23% of cases, with overlap in predicted parameters at the individual patient level.