Key result
Redo-TAVR achieves up to ~95% device success for structural valve deterioration with low coronary risk.
Why the study?
As TAVR expands to younger patients, the likelihood of bioprosthetic valve failure increases, highlighting the need for effective long-term management strategies such as redo-TAVR.
Redo-TAVR is a feasible strategy for managing bioprosthetic valve failure, but requires meticulous preprocedural CT planning to avoid coronary obstruction and patient-prosthesis mismatch.
Answers are grounded in this paper and cite their sources.
Supports redo-TAVR in structural valve deterioration with CT planning; extends observational data but leaves open randomized confirmation.
Transcatheter aortic valve replacement (TAVR) has become the preferred treatment for patients with severe aortic valve stenosis or regurgitation, especially those at intermediate-to-high surgical risk. As TAVR indications expand, the patient population is progressively younger, increasing the likelihood of bioprosthetic valve failure (BVF) over time. This shift highlights the need for effective long-term management strategies. This review focuses on the evolving role of redo-TAVR, specifically the TAV-in-TAV approach, for managing BVF. Key challenges include coronary obstruction, valve migration and device selection. Preprocedural planning, particularly through advanced imaging techniques, such as CT, is critical if complications are to be minimised. This review discusses important technical considerations for TAV-in-TAV procedures, including valve sizing, implantation depth and the impact of valve design on coronary access. A multidisciplinary heart team is essential for assessing patient anatomy and appropriate device selection. Although outcomes appear promising, further research is required to refine procedural techniques and optimise long-term results for patients undergoing redo-TAVR.
No takes yet. Share an insight, caveat, or question.
Fabris et al. (2026) conducted a review in Bioprosthetic valve failure after transcatheter aortic valve replacement. Redo-TAVR (TAV-in-TAV) was evaluated. Redo-TAVR is a viable option for structural valve deterioration, demonstrating a device success rate of 79.7-94.5% and a coronary obstruction rate of less than 1.5%, though careful preprocedural CT planning is essential.