Key result
ANA catheter generates ~6.5-fold higher suction force than standard aspiration catheters despite lower flow.
Answers are grounded in this paper and cite their sources.
BACKGROUND: ANA Advanced Neurovascular Access provides a novel funnel component designed to reduce clot fragmentation and facilitate retrieval in combination with stent-retrievers (SRs) in stroke patients by restricting flow and limiting clot shaving. In previous publications ANA presented excellent in vitro/in vivo efficacy data, especially with fibrin-rich hard clots. We aimed to determine the main physical property responsible for these results, namely suction force versus aspiration flow. METHODS: We evaluated in a bench model the suction force and flow generated by ANA and compared them to other neurovascular catheters combined with a SR (Solitaire). Aspiration flow was evaluated with a flow rate sensor while applying vacuum pressure with a pump. Suction force was determined using a tensile strength testing machine and a purposely designed tool that completely seals the device tip simulating complete occlusion by a hard clot. Suction force was defined as the force needed to separate the device from the clot under aspiration. All experiments were repeated five times, and mean values used for comparisons. RESULTS: Aspiration flow increased with the inner diameter of the device: ANA 1.85±0.04 mL/s, ACE68 3.74±0.05 mL/s, and 8F-Flowgate2 5.96±0.30 mL/s (P<0.001). After introducing the SR, the flow was reduced by an average of 0.57±0.12 mL/s. Due to its larger distal surface, ANA suction force (1.69±0.40 N) was significantly higher than ACE68 (0.26±0.04 N) and 8F-Flowgate2 (0.42±0.06 N) (P<0.001). After introducing the SR, suction force variation was not relevant except for ANA that increased to 2.64±0.41 N. CONCLUSION: Despite lower in vitro aspiration flow, the ANA design showed a substantially higher suction force than other thrombectomy devices.
No takes yet. Share an insight, caveat, or question.
Fernandez-Sanchez et al. (2021) studied Acute ischemic stroke with hard clots (in vitro bench model). ANA Advanced Neurovascular Access catheter (alone and combined with Solitaire stent-retriever) vs. ACE68 and 8F-Flowgate2 neurovascular catheters (alone and combined with Solitaire stent-retriever) was evaluated on Suction force (N) and aspiration flow rate (mL/s) (p=<0.001). The ANA catheter generated higher suction force than ACE68 and 8F-Flowgate2 despite lower aspiration flow.
Synapse has enriched 10 related papers. Consider them for comparative context: