Since May 1994, a new-generation accelerator mass spectrometer (AMS) has been fully operational at the Centre for Isotope Research in Groningen, The Netherlands. The fully automated and high-throughput accelerator mass spectrometry (AMS) system, manufactured by High Voltage Engineering Europa (HVEE) is dedicated to radiocarbon analysis. The HVEE 4130 14 C AMS is able to analyze up to 3000 samples per year. The system is characterized by simultaneous transport of all three isotopes ( 12 C, 13 C, 14 C) and 14 C analysis with a precision below 0.5 pMC and a daily stability below 0.5 pMC. We present here a system description together with stability and performance measurements.