Answers are grounded in this paper and cite their sources.
A digital hygrometer using a polyimide capacitive humidity sensor is developed. The capacitance change of the sensor due to adsorption of water vapor in the atmosphere is detected by a switched-capacitor digital capacitance bridge controlled by a one-chip microcomputer and is displayed as relative humidity with 0.1% resolution. The accuracy of the hygrometer calibrated by a two-point method is solely determined by the temperature dependence and the long-term drift of the dielectric sensitivity of the sensor and is estimated to be 2%.>
No takes yet. Share an insight, caveat, or question.
Matsumoto et al. (1988) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: