Key result
The convolutional neural network model measured left ventricular posterior wall thickness with a relative error of less than 15% compared to hospital measurements, well within the 20% acceptable clinical error.
Why the study?
Manual echocardiographic measurement of left ventricular posterior wall thickness is subjective and difficult to locate edges accurately, introducing errors into the results.
Does a CNN-based automated measurement algorithm accurately measure left ventricular posterior wall thickness compared to manual clinical measurement?
Does a CNN-based automated measurement algorithm accurately measure left ventricular posterior wall thickness compared to manual clinical measurement?
Answers are grounded in this paper and cite their sources.
A CNN-based algorithm can automatically and accurately measure left ventricular posterior wall thickness from echocardiograms with a relative error of less than 15% compared to manual measurements.
BACKGROUND: Clinically, doctors obtain the left ventricular posterior wall thickness (LVPWT) mainly by observing ultrasonic echocardiographic video stream to capture a single frame of images with diagnostic significance, and then mark two key points on both sides of the posterior wall of the left ventricle with their own experience for computer measurement. In the actual measurement, the doctor's selection point is subjective, and difficult to accurately locate the edge, which will bring errors to the measurement results. METHODS: In this paper, a convolutional neural network model of left ventricular posterior wall positioning was built under the TensorFlow framework, and the target region images were obtained after the positioning results were processed by non-local mean filtering and opening operation. Then the edge detection algorithm based on threshold segmentation is used. After the contour was extracted by adjusting the segmentation threshold through prior analysis and the OTSU algorithm, the design algorithm completed the computer selection point measurement of the thickness of the posterior wall of the left ventricle. RESULTS: The proposed method can effectively extract the left ventricular posterior wall contour and measure its thickness. The experimental results show that the relative error between the measurement result and the hospital measurement value is less than 15%, which is less than 20% of the acceptable repeatability error in clinical practice. CONCLUSIONS: Therefore, the measurement method proposed in this paper has the advantages of less manual intervention, and the processing method is reasonable and has practical value.
No takes yet. Share an insight, caveat, or question.
Jian et al. (2020) studied Left ventricular hypertrophy (n=50). Convolutional neural network model for LVPWT measurement vs. Hospital measurement (manual) was evaluated on Relative error between CNN measurement and hospital measurement. The convolutional neural network model measured left ventricular posterior wall thickness with a relative error of less than 15% compared to hospital measurements, well within the 20% acceptable clinical error.