Answers are grounded in this paper and cite their sources.
Optical clearing is an effective method to reduce light scattering of biological tissues that provides significant enhancement of light penetration into the biological tissues making non-invasive diagnosis more feasible. In current report Optical Coherence Tomography (OCT) in conjunction with optical clearing is applied for assessment of deep cartilage layers and cartilage-bone interface. The solution of Iohexol in water has been used as an optical clearing agent. The cartilage-bone boundary becomes visible after 15 min of optical clearing that enabling non-invasive estimation of its roughness: Sa = 10 ± 1 µm. The results show that for 0.9 mm thick cartilage optical clearing is stopped after 50 min with an increase of refractive index from 1.386 ± 0.008 to 1.510 ± 0.009. Current approach enables more reliable detection of arthroscopically inaccessible regions, including cartilage-bone boundary and subchondral bone, and potentially improves accuracy of the osteoarthritis diagnosis.
No takes yet. Share an insight, caveat, or question.
Bykov et al. (2015) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: