An improved variant of the 10mA and 10V dc reference and transfer standard developed for use at CERN is presented. The device achieves very low noise levels at sub-Hz frequencies, which enhances the capability for performance testing of direct-current current transformers (DCCTs) for the High Luminosity Large Hadron Collider project, where good short- and mid-term stability is essential. To characterize the standard, the 10V and 10mA outputs of a single unit were tested simultaneously against two Programmable Josephson Voltage Standards at PTB-Braunschweig. The measurements show negligible low-frequency noise contribution from the internal voltage-to-current conversion, while demonstrating Allan deviation better than 0.01 μA/A on a time scale up to one day, reaching down to few nA/A for shorter measurement times. The operation of multiple paralleled units is also presented as a viable way to further decrease both current and voltage noise.