Answers are grounded in this paper and cite their sources.
Preclinical evaluation demonstrates favorable flexibility and conformability of an adjustable braided stent retriever in a venous sinus model, indicating potential for targeted thrombosis treatment.
Cerebral venous sinus thrombosis (CVST) poses treatment challenges due to fundamental anatomical differences between venous and arterial systems, for which conventional arterial thrombectomy devices are suboptimal. We present a manually adjustable braided NiTi stent retriever designed for CVST that enables real-time control of radial force and thrombus engagement. Finite element analysis demonstrated uniform stress distribution under tensile, compressive, and bending loads. Mechanical tests confirmed low radial force and high flexibility, while deployment in a human venous sinus model showed excellent conformability. This stent overcomes key limitations in venous thrombectomy, providing a promising foundation for CVST-specific devices.
No takes yet. Share an insight, caveat, or question.
Cao et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: