Key result
Lung ultrasound B-lines correlate with right-sided pressures like PASP and PVR, but not PCWP.
Observational (n=100)
Blinded reviewer
Answers are grounded in this paper and cite their sources.
AIMS: Quantification of linear lung ultrasound (LUS) artefacts (B-lines) represents a novel, non-invasive approach to assess pulmonary congestion. We investigated the relationship between the number of B-lines (vertical artefacts arising from the pleural line) and intracardiac pressures. METHODS AND RESULTS: Prior to scheduled right heart catheterization (RHC), 100 subjects underwent LUS of eight zones. A reviewer blinded to the haemodynamic data quantified the number of sonographic B-lines. Of 92 subjects who completed RHC, 79 had adequate LUS data of all zones [median age 61 years, 26 women, median left ventricular ejection fraction (LVEF) 58%, 35 with history of heart failure; 22 postcardiac transplantation]. The number of B-lines correlated with measured right atrial (r = 0.32), pulmonary artery diastolic (PADP) (r = 0.34), mean pulmonary artery (mPAP) (r = 0.43), pulmonary artery systolic (PASP) (r = 0.48) pressures, and pulmonary vascular resistance (PVR) (r = 0.51) (P < 0.005 for all), but not with pulmonary capillary wedge pressure. There was a graded association between tertiles of B-line number and increasing PADP, mPAP, PASP, and PVR (P for trend ≤0.001 for all). Each additional B-line was associated with an increase in PASP of 1 mmHg and an increase in PVR of 0.1 Wood units. These associations remained robust after multivariable adjustment (P = 0.002). Assessment of two inferior lateral zones resulted in similar correlations to the eight-zone method. CONCLUSIONS: Easily obtainable, LUS may be useful in the estimation of right-sided cardiac pressures and PVR. Further evaluation of lung ultrasound as an adjunct to heart failure diagnosis, monitoring, and prognosis is warranted.
No takes yet. Share an insight, caveat, or question.
Platz et al. (2012) conducted an observational in Pulmonary congestion / scheduled right heart catheterization (n=100). Lung ultrasound (LUS) B-line quantification vs. Right heart catheterization haemodynamic pressures was evaluated on Correlation between sonographic B-lines and intracardiac pressures (RAP, PADP, mPAP, PASP) and PVR (r = 0.48 (PASP), r = 0.51 (PVR), p=<0.005). Lung ultrasound B-line number correlated with right-sided cardiac pressures including PASP and PVR, but not with pulmonary capillary wedge pressure.
Synapse has enriched 10 related papers. Consider them for comparative context: