Why the study?
ACCURESP respiratory motion adjustment is recommended during CF and VISITAG Module-guided PVI without in vivo validation, and accurate LAPW RF annotation is crucial to avoid oesophageal thermal injury.
Does the ACCURESP™ 'on' setting improve the accuracy of radiofrequency annotation at the left atrial posterior wall compared to the 'off' setting in patients undergoing pulmonary vein isolation?
Does the ACCURESP™ 'on' setting improve the accuracy of radiofrequency annotation at the left atrial posterior wall compared to the 'off' setting in patients undergoing pulmonary vein isolation?
The ACCURESP™ 'off' setting demonstrated superior catheter movement detection compared to the recommended 'on' setting, providing a more optimal method for annotating stable sites of radiofrequency delivery at the left atrial posterior wall.
Background: During contact force (CF) and VISITAG™ Module (Biosense Webster) guided pulmonary vein isolation (PVI), ACCURESP™ respiratory motion adjustment is recommended, although without in vivo validation. Objective: Since accurate LAPW radiofrequency (RF) annotation is crucial to avoid oesophageal thermal injury, we compared ACCURESP™ setting ("on" versus "off") on RF annotation at the left atrial posterior wall (LAPW). Methods: From a twenty-five patient cohort undergoing CF PVI (continuous RF, 30W) using general anaesthesia and VISITAG™ Module annotation-guidance (force-over-time 100% minimum 1g, 2mm position stability, ACCURESP™ "off"), respiratory motion detection occurred in eight, permitting retrospective comparison of ACCURESP™ settings. Results: There were significant differences in LAPW RF data annotation according to ACCURESP™ setting. Comparing ACCURESP™ "on" versus "off", respectively: Total annotated sites 82 versus 98; Median RF duration per-site 13.3s versus 10.6s (p<0.0001); Median force time integral 177g.s versus 130g.s (p=0.0002); Mean inter-lesion distance (ILD) 6.0mm versus 4.8mm (p=0.002). Considering only annotated site 1-to-2 transitions, three occurred with 0g CF; ACCURESP™ "on" minus "off" difference in RF duration was <0.6s. However, thirteen site 1-to-2 transitions during constant catheter-tissue contact (ILD range 2.1 - 7.0mm) demonstrated a mean difference in annotated RF duration at site 1, of 3.7s (Range: -1.3 - 11.3s). Reconstituted curves displaying catheter position data, CF, impedance and site 1-to-2 transition according to ACCURESP™ setting, demonstrated multiple markers of catheter movement coinciding with ACCURESP™ "off". Conclusions: ACCURESP™ "off" setting demonstrated excellent catheter movement detection properties and represents an optimal method towards the annotation of stable sites of RF delivery at the LAPW.
No takes yet. Share an insight, caveat, or question.
Tomlinson et al. (2019) studied this question.