Why the study?
Does the Kingyield BP210 wrist blood pressure monitor accurately measure blood pressure compared to a mercury sphygmomanometer in adults?
Does the Kingyield BP210 wrist blood pressure monitor accurately measure blood pressure compared to a mercury sphygmomanometer in adults?
The Kingyield BP210 wrist blood pressure monitor meets the accuracy requirements of the European Society of Hypertension-International Protocol for home blood pressure monitoring.
Answers are grounded in this paper and cite their sources.
May support home BP monitoring with the Kingyield BP210; leaves open need for prospective validation in diverse populations.
OBJECTIVE: The present study aimed to evaluate the accuracy of the automated oscillometric wrist blood pressure monitor BP210 for home blood pressure monitoring according to the International Protocol of the European Society of Hypertension. METHODS: Systolic and diastolic blood pressures were sequentially measured in 33 adult Chinese participants (21 women, 51 years of mean age) using a mercury sphygmomanometer (two observers) and the BP210 device (one supervisor). Ninety-nine pairs of comparisons were obtained from 15 participants in phase 1 and a further 18 participants in phase 2 of the validation study. Data analysis was conducted using the ESHIP analyzer. RESULTS: The BP210 device successfully passed phase 1 of the validation study with a number of absolute differences between device and observers within 5, 10, and 15 mmHg for at least 33/45, 44/45, and 44/45 measurements, respectively. The device also achieved the targets for phase 2.1, with 77/99, 95/99, and 97/99 differences within 5, 10, and 15 mmHg, respectively for systolic blood pressure, and with 78/99, 97/99, and 99/99 within 5, 10, and 15 mmHg, respectively for diastolic blood pressure. In phase 2.2, 29 and 25 participants had at least two of the three device-observers differences within 5 mmHg (required≥22) for systolic blood pressure and diastolic blood pressure, respectively. CONCLUSION: The Kingyield wrist blood pressure monitor BP210 has passed the International Protocol requirements, and hence can be recommended for home use in adults.
No takes yet. Share an insight, caveat, or question.
Zeng et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: