Why the study?
Does a novel wristband tonometer provide valid and reliable measurements of central hemodynamics compared to an established pencil-type tonometer in resting participants?
Does a novel wristband tonometer provide valid and reliable measurements of central hemodynamics compared to an established pencil-type tonometer in resting participants?
A novel wristband tonometer provides valid and reliable measurements of central hemodynamics comparable to an established pencil-type tonometer, with faster data acquisition.
Answers are grounded in this paper and cite their sources.
Wristband tonometer may enable easier central BP assessment in practice; leaves open outcome impact and need for prospective validation.
BACKGROUND: Central hemodynamic and augmentation indices are independent predictors of cardiovascular events and all-cause mortality that can be estimated noninvasively by pulse wave analysis. The purpose of this study was to assess the reliability and validity of a newly engineered wristband tonometer for acquiring radial artery pressure waveforms. METHODS: Radial artery pulse pressure waveforms were evaluated with an established pencil-type and a novel wristband tonometer in 31 participants (aged 30.2±9.5 years) resting in a supine position. Pulse wave analysis was executed using the same validated generalized transfer function (SphygmoCor) for both tonometers. RESULTS: A significant difference in time to data acquisition between tonometers was observed (-70.2±147.7 s; P < 0.05; wristband faster). The wristband tonometer had significantly lower within-subject coefficients of variation (CV) compared with the pencil-type tonometer in aortic pulse wave height (-2.67% ± 5.51%; P < 0.05) and time to reflection (-2.26% ± 6.16%; P < 0.01). No other differences in CV were observed. Slight but statistically significant mean differences between tonometers were observed in aortic systolic blood pressure (ASBP; 0.43±1.08 mm Hg; P < 0.05; wristband lower), aortic pulse pressure (APP; 0.43±0.96 mm Hg; P < 0.05; wristband lower), and round-trip travel time of the reflected pressure wave (Δtp; 3.58±12.86 ms; P < 0.05; wristband higher). However, ASBP, APP, and Δtp measurements were highly correlated (r = 0.9970, r = 0.9953, and r = 0.8838, respectively, P <0.0001) between tonometers; within-subject and between tonometer significant mean differences were within clinical ranges. CONCLUSIONS: This novel, hands-free platform may be interchangeable with the commonly used pencil-type tonometer, heralding new directions in noninvasive in vivo vascular research and clinical application.
No takes yet. Share an insight, caveat, or question.
Beck et al. (2014) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: